Suntex
Inductive Conductivity Transmitter
Inductive conductivity transmitter with an anti-fouling electrodeless sensor and range up to 2000 mS...
Intelligent conductivity/resistivity transmitter compatible with 2- or 4-electrode cells, with automatic temperature compensation and isolated 4~20mA output. IP65 rated for RO systems, pure water, boiler water, cooling water and water treatment.
| Models | EC-4110 EC-4110-RS |
|---|---|
| Measuring Modes | Conductivity / Resistivity / Temperature |
| Conductivity Range | 0.00 µS/cm to 200.0 mS/cm (5 ranges, auto or fixed) |
| Resistivity Range | 0.00 to 20.00 MΩ·cm |
| Temperature Range | -30.0 to 130.0°C |
| Resolution | 0.01 / 0.1 / 1 µS/cm, 0.01 / 0.1 mS/cm (Conductivity) 0.01 MΩ·cm (Resistivity) 0.1°C (Temp) |
| Accuracy | ±1% (Conductivity) ±1% (Resistivity) ±0.2°C (Temp) |
| Cell Constant | 0.01, 0.05, 0.1, 0.5, 10.0 cm⁻¹ fixed Freely selectable 0.008 to 19.99 cm⁻¹ |
| Calibration | Automatic with 3 built-in standard solutions Manual cell constant adjustment |
| Temperature Compensation | Automatic (NTC 30KΩ or Pt-1000) or manual |
| Temperature Coefficient | Linear 0.00% to 40.00% Non-linear for natural water |
| Analog Output 1 | Isolated 0/4~20mA for measuring range, max load 500Ω |
| Analog Output 2 | Isolated 0/4~20mA for temperature, max load 500Ω (EC-4110 only) |
| Serial Interface | RS-485 Modbus RTU/ASCII (RS model only) |
| Logbook | 50 event records (RS model only) |
| Relay Contact | 240VAC, 0.5A max |
| Power Supply | 100 to 240 VAC ±10%, 50/60 Hz, 7W max |
| Ambient Temperature | 0 to 50°C |
| Installation | Panel, wall or pipe mounting |
| Dimensions | 96 x 96 x 132 mm |
| Panel Cut-out | 93 x 93 mm |
| Weight | 0.5 kg |
| Protection | IP65 (NEMA 4X) |